Edge Ultrasound System User's Guide Page 152

  • Download
  • Add to my manuals
  • Print
  • Page
    / 198
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 151
140
(a)This index is not required for this operating mode; value is <1.
(b)This transducer is not intended for transcranial or neonatal cephalic uses.
# No data are reported for this operating condition since the global maximum index value is not reported for the reason
listed. (Reference Global Maximum Index Value line.)
Data are not applicable for this transducer/mode.
Table 17: Transducer Model:
HFL50x
Operating Mode:
CPD/Color
Index Label M.I.
TIS TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
1A
aprt
>1
Global Maximum Index Value 1.3 (a) (b)
Associated Acoustic
Parameter
p
r.3
(MPa) 3.05
W
0
(mW) #— —#
min of
[W
.3
(z
1
),I
TA.3
(z
1
)]
(mW)
z
1
(cm)
z
bp
(cm)
z
sp
(cm) 1.2
d
eq
(z
sp
) (cm)
f
c
(MHz) 5.36 # #
Dim of A
aprt
X(cm) #— #
Y(cm)
#— #
Other Information
PD (μsec) 0.521
PRF (Hz) 8233
p
r
@PII
max
(MPa) 3.81
d
eq
@Pll
max
(cm)
Focal Length FL
x
(cm) #— #
FL
y
(cm) #— #
I
PA.3
@MI
max
(W/cm
2
)
494
Operating
Control
Conditions
Control 1: Mode Any
Control 2: Exam Type Any
Control 3: Optimization/Depth Low/3.3
Control 4: PRF Any
Page view 151